Introduction
Most industrial vacuum processes span six decades of pressure—from atmosphere during venting and roughing to high vacuum during deposition, etching, or analysis. A single gauge cannot deliver accurate, continuous readings across this entire range. Thermal-conductivity gauges lose sensitivity below 10−3 Torr; ionization gauges cannot operate reliably near atmosphere. The practical solution is to combine a Pirani vacuum transmitter with a cold cathode gauge, creating a seamless full-range measurement system that engineers and procurement teams trust for repeatability, uptime, and simplified inventory.
At Poseidon Scientific, we designed the VG-SP205 Pirani Vacuum Transmitter and the VG-SM225 Cold Cathode Vacuum Gauge to work together as a matched pair. Both share KF25 flanges, compact footprints, and straightforward electrical interfaces, so integration is fast and cost-effective. This article explains the need for full-range coverage, the natural overlap between the two technologies, proven switching strategies, automation logic, a real-world industrial example, and the specific reasons the VG-SP205 + VG-SM225 combination outperforms mixed-vendor solutions.
The Need for Full-Range Vacuum Coverage
Vacuum processes rarely stay in one pressure decade. A semiconductor load lock starts at atmosphere, drops to 10−3 Torr during roughing, then reaches 10−6 Torr or lower before wafer transfer. A PVD coater requires stable high vacuum for sputtering but must vent and re-pump between runs. A vacuum furnace cycles between rough vacuum for initial pump-down and high vacuum for annealing.
Using only a Pirani leaves you blind below 10−3 Torr. Using only a cold cathode forces you to guess during roughing and risks sensor damage from sudden pressure spikes. Full-range coverage eliminates these gaps, enables automatic pump-down sequencing, provides continuous data for trending and alarms, and simplifies compliance with ISO or SEMI standards. A properly paired system delivers one clean pressure signal to the PLC or HMI from atmosphere all the way to 10−7 Torr—without manual intervention or extra calibration steps.
Overlapping Measurement Zones
The VG-SP205 and VG-SM225 are intentionally engineered with a precise overlap that makes handoff seamless:
- VG-SP205 Pirani: 1 × 10−3 to 760 Torr (air). Peak accuracy (±15 %) occurs in the 10−2–10−1 Torr decade.
- VG-SM225 Cold Cathode: 1 × 10−7 to 1 × 10−3 Torr (air). Stable readings continue up to 10−3 Torr with ±20 % accuracy.
The crossover point at approximately 10−3 Torr is where the Pirani’s thermal-conductivity signal begins to roll off and the cold cathode’s ion-current signal becomes dominant. This overlap is wide enough (nearly one full decade) for the controller to compare both readings and switch cleanly without pressure jumps or dead bands. The result is continuous, traceable data across six decades using only two compact transmitters.
Switching Strategy
Three practical switching methods are commonly used, all supported by the Poseidon pair:
- Simple threshold switch: The PLC monitors the Pirani until pressure falls below 5 × 10−3 Torr, then disables the Pirani signal and enables the cold cathode. Hysteresis (e.g., switch back at 2 × 10−3 Torr) prevents rapid toggling.
- Blended signal: In the overlap zone both gauges are read simultaneously; the controller applies a linear or weighted average that smoothly transitions from Pirani-dominant to cold-cathode-dominant output. This eliminates any discontinuity on trend charts.
- Auto-select by pressure stability: The system compares repeatability of both signals and automatically chooses the more stable reading. This method is ideal for processes with variable gas loads.
All three strategies require no additional hardware. The VG-SP205’s RS232 output and the VG-SM225’s logarithmic 0–10 V analog output are directly compatible with any modern PLC or data-acquisition system.
Automation Logic for Seamless Operation
Typical PLC logic for the paired gauges is straightforward and reusable:
- Power both transmitters continuously (VG-SP205 at 5 V, VG-SM225 at 24 V).
- Read the Pirani pressure via RS232 and the cold cathode via analog input.
- Below 10−3 Torr, use the cold cathode value for control and trending; above 5 × 10−3 Torr, use the Pirani.
- In the overlap band, blend the signals or apply the chosen switching rule.
- Monitor status bits: the VG-SP205 flags communication or sensor faults; the VG-SM225’s analog voltage drops below 2 V on over-range or ignition failure.
- Add interlocks: prevent high-voltage enable on the cold cathode until the Pirani confirms pressure <1 Torr; inhibit roughing-pump start if either gauge reads atmosphere with the chamber door open.
This logic runs in a single function block and can be dropped into any new project. Sample code for Siemens, Allen-Bradley, and Modbus is available on both product pages. Engineers report the pairing reduces pump-down time by 20–30 % and eliminates false alarms caused by range limitations.
Industrial Example: PVD Coating System
A production PVD line for optical coatings illustrates the pairing in action. The load lock and process chamber each receive one VG-SP205 (foreline monitoring) and one VG-SM225 (chamber base pressure). The PLC:
- Uses the Pirani to confirm safe roughing and triggers the turbo pump once pressure drops below 1 Torr.
- Switches to the cold cathode at 10−3 Torr for precise high-vacuum control during reactive sputtering.
- Logs a continuous pressure curve for every batch, enabling predictive maintenance when base pressure drift exceeds 10 %.
Result: coating thickness uniformity improved from ±8 % to ±2 %, scrap rate dropped 40 %, and calibration frequency was reduced from quarterly to annual. The same logic now runs on six identical coaters with one spare-parts kit serving both gauge types.
Why the VG-SP205 + VG-SM225 Pairing Works So Well
Other combinations require adapters, different flanges, mismatched power supplies, or complex scaling. The Poseidon pair was purpose-designed to eliminate those issues:
- Identical KF25 flanges and mounting dimensions simplify mechanical integration.
- Low power draw (<2 W Pirani, <7 W cold cathode) and shared 24 V / 5 V supplies reduce cabinet space.
- RS232 (Pirani) + logarithmic analog (cold cathode) outputs match any controller without converters.
- Built-in status LEDs and error flags give instant visual and digital diagnostics.
- Compact size (<80 mm long) and bake-out rating to 150 °C fit laboratory and production chambers alike.
- Proven repeatability (±5 % Pirani, ±20 % cold cathode) and 12-month calibration intervals lower total cost of ownership.
Because both gauges were developed together, spare parts, cables, and documentation are unified—procurement teams order one kit instead of two separate vendor packages. The result is faster commissioning, lower inventory, and fewer integration surprises.
Conclusion
Combining a Pirani and cold cathode gauge is the most practical way to achieve accurate, continuous vacuum measurement from atmosphere to high vacuum. The natural overlap at 10−3 Torr, simple switching strategies, and straightforward automation logic make the transition invisible to the process. The Poseidon VG-SP205 and VG-SM225 pairing delivers this capability in the smallest footprint with the easiest integration and lowest long-term cost.
Ready to implement full-range vacuum monitoring in your system? Our applications team specializes in helping engineers and procurement specialists select the right combination, review chamber layouts, provide sample PLC code, and deliver rapid quotations. Contact us today for a no-obligation technical consultation—simply visit the product pages below or reply to this article.
VG-SP205 Pirani Vacuum Transmitter
VG-SM225 Cold Cathode Vacuum Gauge
At Poseidon Scientific we design vacuum instrumentation that works together—giving you seamless coverage, simplified integration, and the confidence that your pressure data is always accurate and ready for automation.



